1523/1524 Reference Thermometers

This handheld digital Reference Thermometer from Fluke is a precision, portable, electronic thermometer with trending and logging capability for NIST-traceable calibratin and high accuracy. The 1523 reads a PRT, RTD, thermocouple or thermistor and the 1524 can read two probes with simultaneous measurements.

Measure, graph and record three sensor types with one tool

1523 and 1524 Handheld Reference ThermometersHigh accuracy

  • PRTs: up to ± 0.011 °C
  • Thermocouples: up to ± 0.24 °C
  • Precision thermistors: ± 0.002 °C

Two models

  • 1523: Single channel standard model with memory for 25 readings
  • 1524: Two channels; memory for logging 15,000 measurements; real-time clock for time and date stamps

Finally, a reference thermometer as versatile as you are

The new 1523/24 Reference Thermometers from Fluke’s Hart Scientific division measure, graph, and record PRTs, thermocouples, and thermistors. These new thermometer readouts deliver high accuracy, wide measurement range, logging, and trending, all in a handheld tool you can take anywhere.

The 1523/24 lets you handle field applications, laboratory measurements, and data logging with ease. And with the dual channel measurement capabilities of the model 1524, you can do twice the work in half the time.

Make accurate, consistent measurement...anywhere

You need accuracy for compliance, product yields, energy savings, and consistent results. The 1523/24 uses current reversal, a technique used in high-end instruments that eliminates thermal EMFs, for precision temperature measurements. Specifications are guaranteed from –10 °C to 60 °C ambient. Special precision resistors and a highly stable reference voltage source keep 1523/24 accuracy virtually insensitive to environmental temperature.

Like all Fluke handheld tools, the 1523/24 Reference Thermometers endure rigorous testing in temperature extremes and under harsh conditions of vibration, so you can take it with confidence anywhere you need to go. An optional magnetic hanger allows you to hang the thermometer for easy viewing while freeing your hands to focus on the job.

Monitor trends in the lab or in the field 1523 Thermometer used in the field to calibrate temperature sensors.

See trends graphically on the 1523/24 thermometer’s 128x64-backlit LCD display. You can change the graph’s resolution at the touch of a button. Now it is easy to see when the temperature is stable (without statistics or long delays) or to monitor processes over time to verify correct operation.

Hold readings on the display at the touch of a button, or document up to 25 readings and associated statistics for easy retrieval. Statistics include the average maximum and minimum values, and the standard deviation. View them through the display or by uploading it to a PC via RS-232 connection and 9940 I/O Took Kit software, included free. To monitor and log more data over time, use a PC and optional LogWare II software.

RS-232-to-USB adapters are available for those who prefer USB connectors. Battery power lasts at least 20 hours on three AA batteries, or use the dc power adapter for extended periods of measurement. Power saving features can be enabled or disabled for longer battery life or greater convenience.

INFO-CON connectors ensure correct temperature conversion

Inside the INFO-CON, a memory chip keeps calibration information for the attached probe. Simply plugging in the probe uploads the information to the readout. The connector transfers this information to the 1523/24 automatically, ensuring the correct temperature conversion for accurate, hassle-free measurements.

Probes may be locked by password to specific channels and readouts for security or for system calibration traceability. Plug any thermocouple with mini-thermocouple jacks into an optional universal thermocouple adapter for convenient measurement. Each thermocouple adapter or standard connector supports reference junction compensation (RJC) with its own internal precision thermistor.

Two models let you make the best choice for your application

1523 single-channel thermometer

1524 two-channel thermometer

1523 single-channel thermometer 1524 two-channel thermometer
1 RS-232 serial interface connector. For PC communications, uploading and downloading data from memory and from the probe INFO-CON connectors. 1 RS-232 serial interface connector
2 Sensor connector (PRT, thermocouple, or thermistor) 2 Channel 1 sensor connector (PRT, thermocouple, and thermistor)
4 External power adapter connection, for continuous use without changing batteries. Alternatively, 3 AA batteries will last more than 20 hours in the field. 3 Channel 2 sensor connector (PRT, and thermistor)
  4 External power adapter connection
The 1523 Reference Thermometers are versatile single-channel thermometers that measure, graph and record three sensor types with one tool. Support for PRTs/RTDs, thermocouples, and thermistors provide flexibility to choose the right probe for the job. The new 1524 Reference Thermometers help you do twice the work in half the time. Two channels and three sensor types and high-speed measurement make you more productive and make model 1524 the one reference thermometer you need to own. It has all the features of the 1523, and it’s a data logger too. A real-time clock and memory for 15,000 time and date stamped measurements mean everything you are going to need is in this package. Log up to three times per second, or once every hour or other options in between. Download the data to a PC for analysis when you need it.

 Specifications

1523

1524

Input channels

1

2

Resolution

PRTs and thermistors: 0.001°
thermocouples: 0.01°

Logging

25 readings with statistics

25 readings with
statistics15,000 time
and date stamped

Sample interval (normal)

1 second

1 second
(simultaneous measurement)

Typical sample interval (fast mode)*

0.3 seconds

Sensor types

PRTs, RTDs, thermistors, and thermocouples

Thermocouple types

C, E, J, K, L, M, N, T, U, B, R, S

Statistics

Maximum, minimum, average, standard deviation

Trending

Scale: ± 10 °C (18 °F), ± 1 °C (1.8 °F), ± 0.1°C (0.18 °F),
± 0.01°C (0.018 °F), 10 minutes of real-time data

Power requirements

3 AA alkaline batteries, 12 V dc universal power supply

Size (HxWxD)

96 mm x 200 mm x 47 mm (3.75 in x 7.9 in x 1.86 in)

Weight

0.65 kg (1.4 lb)

Computer interface

RS-232, 9940 I/O ToolKit software included

Safety

EN61010-1:2001, CAN/CSA C22.2 No. 61010.1-04

Environmental conditions for best accuracy: 13 °C to 33 °C (55.4 °F to 91.4 °F)

Millivolt range and accuracy

–10 mV to 75 mV, ± (0.005 % + 5 µV)

Internal reference junction compensation

± 0.2 °C (± 0.36 °F)

Resistance range and accuracy

0 W to 400 W ± (0.004 % + 0.002 W)
200 W to 50 kW ± (0.01 % + 0.5 W)
50 kW to 500 kW ± (0.03 %)

Temperature coefficient, voltage:
–10 °C to 13 °C , +33 °C to 60 °C
(14 °F to 55.4 °F, 91.4 °F to 140 °F)

± (0.001 %/°C + 1 mV/°C)

Temperature coefficient, resistance:
–10 °C to 13 °C , +33 °C to 60 °C
(14 °F to 55.4 °F, 91.4 °F to 140 °F)

0.0008 %/°C + 0.0004 W (0 W to 400 W)
0.002 %/°C + 0.1 W (0 W to 50 kW)
0.06 %/°C + 0.1 W (50 kW to 500 kW)

Excitation current, resistance

1 mA (0 W to 400 W)
10 µA (0 W  to 50 kW)
2 µA (50 kW to 500 kW)

*See technical manual for sample interval details by sensor type and number of inputs.

Equivalent temperature accuracies for selected sensors derived from primary specifications (Ω, mV)

Temperature Thermocouples

Type

Range

Measure accuracies

K

–200 °C to 0 °C
(-328 °F to 32 °F)

± 0.61 °C
(± 1.10 °F)

0 °C to 1370 °C
(32 °F to 2498 °F)

± 0.24 °C
(± 0.43 °F)

R

–20 °C to 0 °C
(4 °F to 32 °F)

± 1.09 °C
(± 1.96 °F)

0 °C to 500 °C
(32 °F to 932 °F)

± 0.97 °C
(± 1.71 °F)

500 °C to 1750 °C
(932 °F to 3182 °F)

± 0.49 °C
(± 0.88 °F)

S

–20 °C to 0 °C
(4 °F to 32 °F)

± 1.05 °C
(± 1.89 °F)

0 °C to 500 °C
(32 °F to 932 °F)

± 0.95 °C
(± 1.71°F)

500 °C to 1750 °C
(932 °F to 3182 °F)

± 0.56 °C
(± 1.01 °F)

Accuracies are based on internal reference junction compensation. Refer to technical manual for greatly improved accuracies using external reference junctions.

Temperature, RTD Ranges, and Accuracies (ITS-90)

Accuracy ± °C 4 (K=2)

± 0.011 at –100 °C

± 0.015 at 0 °C

± 0.019 at 100 °C

± 0.023 at 200 °C

± 0.031 at 400 °C

± 0.039 at 600 °C

Resolution: 0.001°C (0.001°F)

Temperature, Thermistor

Accuracy ± °C (K=2)

± 0.002 at 0 °C

± 0.003 at 25 °C

± 0.006 at 50 °C

± 0.014 at 75 °C

± 0.030 at 100 °C

Resolution: 0.001 °C (0.001 °F)

Based on a 10 kW (at 25 °C) thermister with a beta value of 4000 W. See technical manual for details.

Accuracies of selected readout/probe
combinations (±°C) (K=2)

Temperature

5616-12

5615-6

5627A-12

5610-9

-200 °C (-328 °F)

0.014

0.025

0.027

n/a

0 °C (32 °F)

0.021

0.021

0.049

0.009

100 °C (212 °F)

0.027

0.028

0.065

0.030

300 °C (572 °F)

0.040

0.043

0.103

n/a

420 °C (788 °F)

0.050

n/a

0.130

n/a

Includes readout accuracy,
probe calibration, and probe drift

 

Ordering Information

1523* Reference Thermometer, Handheld, 1 Channel
1524* Reference Thermometer, Handheld, 2 Channel, Data Logger
1523-P1 1523 Bundled with 5616 PRT [–200 °C to 420 °C (-328 °F to 788 °F), NIST Traceable Probe Calibration, 100 ohm, 6.35 mm x 305 mm (1/4 in x 12 in)],
Universal TC INFO-CON Connector, TPAK, and Case
1523-P2 1523 Bundled with 5628 PRT [–200 °C to 660 °C (-328 °F to 1220 °F), Accredited Probe Calibration, 25 ohm, 6.35 mm x 305 mm (1/4 in x 12 in)],
Universal TC INFO-CON Connector, TPAK, and Case
1523-P3 1523 Bundled with 5627A PRT [–200 °C to 420 °C (-328 °F to 788 °F), Accredited Probe Calibration, 100 ohm, 6.35 mm x 305 mm (1/4 in x 12 in)],
Universal TC INFO-CON Connector, TPAK, and Case
1524-P1 1524 Bundled, with 5616 PRT, Universal TC INFO-CON Connector, TPAK, and Case
1524-P2 1524 Bundled with 5628 PRT, Universal TC INFO-CON Connector, TPAK, and Case
1524-P3 1524 Bundled with 5627A PRT, Universal TC INFO-CON Connector, TPAK, and Case
*Requires an optional probe

Calibration options

1523-CAL 1523 Accredited Calibration
1524-CAL 1524 Accredited Calibration
1929-2 System Verification, PRT with Readout, Accredited
1929-5 System Verification, Thermistor with Readout, Accredited
1930 System Calibration, PRT with Readout, Accredited
1935 System Calibration, Thermistor with Readout, Accredited
1925-A Accredited Thermistor Calibration, 0 °C to 100 °C (23 °F to 212 °F)

Included accessories

NIST traceable certificate of calibration, users guide, CD-ROM (contains technical manual), 12 V dc universal power supply, RS-232 cable, 9940 I/O ToolKit software

Optional accessories

5610-9-P Probe, Precision Thermistor, Stainless Steel, 3.18 mm x 228.6 mm (1/8 in x 9 in),
0 °C to 100 °C (32 °F to 212 °F), NIST traceable calibration
5615-6-P Probe, PRT, 100 ohm, 4.76 mm x 152.4 mm (3/16 in x 6 in), –200 °C to 300 °C
(–328 °F to 572 °F), Accredited Calibration
5609-9BND-P Probe, PRT, 100 ohm, 6.35 mm x 381 mm (1/4 in x 15 in), 90° bend at 9 inches,
–200 °C to 660 °C (–328 °F to 1220 °F), Requires Calibration (i.e. 1924-4-7)
FLK80P1 80PK-1, Probe and adapter (2373-LTC), Thermocouple, Beaded Type K, –40 to 260 °C
FLK80P3 80PK-3A, Probe and adapter (2373-LTC), Thermocouple, Surface Measurement Type K, 0 to 260 °C
9935-S  Software, LogWare II, Single User
1523-CASE Case, 1523/1524 Readout and Probe Carrying Case
FLUKETPAK TPAK, Meter Hanging Kit   TPAk
2373-LPRT Adapter, Lemo to Mini Grabbers (4-wire) Adapter
2373-LTC Adapter, Lemo to Universal TC (TC) TC Adapter
2373-L21 Adapter, Lemo to INFO-CON (allows the 1523/1524 to read probes terminated with the 1521/1522 INFO-CON)
2384-P INFO-CON Connector, PRT (Gray Cap), Spare
2384-T INFO-CON Connector, TC (Blue Cap), Spare
2525-156 Isolator, RS-232 Optical 12 V dc (1523/1524), 115 V
2523-256 Isolator, RS-232 Optical 12 V dc (1523/1524), 230 V

Downloads

The following items are available for download in Adobe Acrobat (.pdf) format and require Adobe Reader or Acrobat. Click on the link to view in your browser or right click and “Save Target As...” to download to your computer.

Note: Users Guides are available for download in separate language versions. Visit our manual page or go directly to our ftp site area - ftp.hartscientific.com/manuals

Use our feedback form for more information.

(return to top)

 

 

 

Send mail to webmaster@hartscientific.com with questions or comments about this web site.
Copyright © 1998-2009 Fluke Corporation